PNGE 470 Colorado State University Accuracy of Positive Displacement Meters Lab Report Hi there I attached the instructions for the lab and 2 samples work,

PNGE 470 Colorado State University Accuracy of Positive Displacement Meters Lab Report Hi there I attached the instructions for the lab and 2 samples work, I have one more sample also if u want PNGE 470
EXPERIMENT 2
SONIC NOZZLE FLOW PROVER
INTRODUCTION
The sonic nozzle flow prover is a portable meter proving device which can be
used to test positive displacement meters. It is operable at a pressure of 5
psig or more by passing gas or air through the meter and the flow prover and
discharging it into the atmosphere. The minimum test pressure is governed
by the fact that sufficient pressure must exist across the nozzle to insure sonic
velocity at the nozzle throat. This requirement is met when the nozzle
discharge absolute pressure is less than 85% of the absolute nozzle upstream
pressure (80% for nozzle sizes A and B). Approximate critical flow rates of
sonic flow nozzles are given in table 1 below. Table 1 can be used to
determine the nozzle sizes which should be used for the tests.
TABLE 1
Size Code
A
B
C
D
E
F
G
H
J
Throat Size
Inch
0.094
0.125
0.183
0.250
0.312
0.375
0.438
0.500
0.625
Time for One ft3 of Air
Seconds
36.00
18.450
8.00
4.598
2.88
2.0355
1.47
1.1404
7.290
PNGE 470
EXPERIMENT 2
TEST PROCEDURE
SONIC NOZZLE FLOW PROVER
1. Attach the Prover close to the outlet of the meter (See Figure 1).
2. Pass air through the meter and the prover. The air must be passed
through the prover until the temperature on the thermometer become
constant. The pressure should be maintained throughout the test.
3. Use a stop watch to record the time (in seconds) required for a specific
number of cubic meter to register on the meter.
4. It is recommended to conduct the test at three flow rates
(Approximately 20%, 50%, and 80% of the meter capacity). Do not
exceed the maximum capacity of the meter.
5. The test time should not be less than 100 seconds, preferably 150 to
200 seconds.
6. Calculate the Proof of the meter using:
? t ?? 1 ?? F T P F ?
%Proof = ? ?? ?? a m n R ?
? ts ?? Vt ?? Tn Pm ?
t = Test Time, seconds
ts = Standard Time, seconds ( Stamped on the Nozzle )
Vt = Test Volume, cu.ft.
Fa = Factor for Air, Table 2
Tm = Meter Temperature, oR
Tn = Nozzle Temperature, oR
Pm = Meter Pressure, psia
Tn = Nozzle Pressure, psia
FR = Reynold’s Number Factor, Table 3
When “%Proof” is more than 100%, the meter is slow.
PNGE 470
EXPERIMENT 2
SONIC NOZZLE FLOW PROVER
FIGURE 1. PROVER ASSEMBLY
PRECAUTIONS
1. Noise level will be high and earplugs must be used.
2. The nozzles should be handled carefully.
3. Errors are introduced if pressure and temperature are not constant
during the test.
TABLE 2
TABLE 3.FR Factor
6/30/2020
PNGE 470
Lab 2
mohammed Alfaqeer, Raeed Alzyadi
Objective:
The main goal of the experiment is to show the accuracy of the gas meter by comparing a known
volume of air with the gas register and calculating the percentage proof of the meter.
Equipment:
1- Different sizes for the sonic flow nozzle prover.
2- Valve to control the flow.
3- Pressure tap.
4- Thermometer.
5- Stop watch.
Theory:
The sonic nozzle flow prover dates back to over 25 years and has been under constant
improvement over the years and became very important with the gas production. It is a portable
meter device that is used to test the positive displacement and is operable at pressures of 5 psig
or higher. Where the gas would flow through the meter and the prover then discharged into the
atmosphere.
Moreover, enough pressure must exit across the nozzle to achieve the sonic velocity at
the nozzle throat. This requirement is achieved when the discharge pressure is less than 85% of
the upstream pressure, or 80% for nozzles A and B.
The percentage of proof is used to determine if the meter is considered slow or fast.
%?????????? =
?? 1 ???? ? ???? ? ???? ? ????
? ?
???? ????
???? ? ????
t: test time in seconds
ts: standard time (on the nozzle) in seconds
Vt: test volume in ft3
Fa: factor of air from tables
Tm: meter temperature in Rankine
Tn: Nozzle temperature in Rankine
Pm: meter pressure in psia
Pn: Nozzle pressure in psia
FR: Reynolds number factor from tables
Note that when %?????????? > 100%, it is considered slow. If %?????????? < 100%, it is fast. Procedure & Precaution: 1- Using earplugs since the level of the noise is high. 2- Handling the nozzles carefully for safety reasons. 3- Connecting the prover to the O ring. 4- Connecting the nozzle flow to the meter outlet. 5- Passing the air through the prover until the temperature on the thermometer becomes constant. 6- Recording the time in seconds for each size. 7- The time is preferred to be between 150 to 200 seconds. 8- Using the proper equations, calculating the proof of the meter. Results: B Test time Nozzle pressure nozzle temperature meter pressure Meter Temperature Standard time t Pn Tn Pm Tm ts 177.70 29.73 537.00 30.73 537.00 18.45 sec psia F psia F sec Volume air factor Reynold factor Percentage proof Percent Accuracy Vt Fa Fr %Proof %Accuracy 10.00 ft^3 101.62 1.001 94.78 105.50 Test time Nozzle pressure nozzle temperature meter pressure Meter Temperature Standard time Volume air factor Reynold factor Percentage proof Percent Accuracy D t Pn Tn Pm Tm ts Vt Fa Fr Proof %Accuracy 180.58 28.23 537.00 29.23 537.00 4.60 40.00 101.62 1.000 96.36 103.78 sec psia F psia F sec ft^3 199.48 26.73 537.00 27.73 537.00 2.04 100.00 101.62 1.000 96.00 104.17 sec psia F psia F sec ft^3 F Test time Nozzle pressure nozzle temperature meter pressure Meter Temperature Standard time Volume air factor Reynold factor Percentage proof Percent Accuracy t Pn Tn Pm Tm ts Vt Fa Fr Proof %Accuracy Discussion of results: From the tables above, we can see that the highest percentage of proof was for size code D while the lowest percentage of proof was for the size code B. Moreover, since size codes B, D and F all show a percentage of proof less than 100%, they are all considered to be fast. In addition, it can be noted that whenever the throat size increases, the time it takes for the air volume to be registered by the meter increases. So it can be established that there is a relationship between them. Furthermore, the upstream pressure decreases as the throat size increases. Indicating an inversely proportional relationship between them. Conclusion: Appendix: Sample calculation for Size case B: ???? = 15 + 14.73 = 29.73 ???????? ???? = 29.73 + 1 = 30.73 ???????? ???? = 77 + 460 = 537 ?? = ???? For finding Fa and Fr, interpolation is used: ????: 78 ? 76 78 ? 77 = , ?????????????? ?????? ???? ???? ?????? ???? = 101.62 101.71 ? 101.52 101.71 ? ???? ???? : 20 ? 10 20 ? 15 = , ?????????????? ?????? ???? ???? ?????? ???? = 1.001 1.002 ? 1.000 1.002 ? ???? %?????????? = 177.7 1 101.62 ? 537 ? 29.73 ? 1.001 ? ? = 94.78% 18.45 10 537 ? 30.73 Summer 2020 PNGE470 PNGE 470 Lab#2 7/1/2020 Objective: Summer 2020 PNGE470 The main goal for this experiment is to conduct an experiment using the sonic nozzle flow prover device. This device is used to test positive displacement meters over a certain time. Also, it is operable at 5 psig or more and its main component is passing fluid through the meter and the flow prover orifice and discharging it into the atmosphere. For this experiment, the sonic nozzle is to be tested at the required pressure. Three different runs for this data with three different flow rates. Another aspect for this experiment is to ensure that sufficient pressure exist through the nozzle. Also, when the nozzle is discharged the absolute pressure is less than 85% of the absolute nozzle. Table1 can show the approximate critical flow and the time taken for different types of nozzles. Summer 2020 PNGE470 Equipments: Thermometer Pressure tap Flow Sonic flow Used to measure Temprature Used to isolate pressure Flow in (downstream) nozzle Flow out (Upstream) = atmosphere Three different types of sonic flow nozzles is to be used in this experiment. Theory: A prover is a device used to measure the accuracy of a gas meter and checks weather it meets the standard conditions. There are many types of nozzles prover. However, the sonic nozzle prover was used to conduct this experiment. Sonic nozzle prover is a device that is used to test positive displacement meters over a certain time. Positive displacement meters measure the volume of gas and or liquid that passes through and to attain the flow rate. The sonic nozzle prover runs at pressure of 5 psig or more. Also, sufficient amount of pressure must exist through the nozzle in order to maintain constant velocity. This experiment can be done by passing certain amount of cubic meters of a fluid through the orifice of the sonic nozzle prover and will be released into the atmosphere. Different throat sizes have different Summer 2020 PNGE470 index rates and time for one cubic foot of air to flow. The proof meter stays constant despit the change in operating pressure. The proof can be calculated as a percentage if the amount of gas or liquid that passes through the orifice is known using the equation below. %?????????? = ???????? ???????? ? 100 ?????????????? ???????? This equation illustrate the relationship between the test time and the correct time both of which are measured in seconds. Also if the %proof is more that 100%, it means that the meter is slow. Test time is for the predetermine volume to register in the meter and the correct time is for the time that is on the nozzles. Correct time might be different for each type of nozzle. Table1 shows the type of nozzle and its associated data for a one cubic foot of air to flow. However, it is different when dealing with more than one cubic foot of dry gas and a new equation has been found and this equation is showen below. ?? 1 ???????????????? %?????????? = ( )( )( ) ???? ???? ???????? Where: t = test time, seconds ts = Standard time, seconds(on the nozzle) Vt = test volume(cu.ft) Fa = factor for air Tm = meter temp Pn= nozzle pressure Fr= rynold`s number factor Tn= nozzle temp pm= meter pressure Procedure: 1- Attach the prover close to the outlet to the out let of the meter. A tight connection is important to avoid pressure drop between the flow meter and the prover. Summer 2020 PNGE470 2- The valve is opened so the gas can pass through the meter and the prover and the time should be measured using a stop watch. The gas should keep flowing through the prover and the meter until the thermometer became constant. 3- The rate of time is being conducted for three different flow rates each rate shold be measured at an index rate based on the type of nozzle. These index ranges from 10% and 20% to 50%, and from 80% to 100% and are rated for the capacity of the meter. 4- The time for each trial should not be less than 100 seconds, preferably from 150 second up to 200 seconds. 5- The time stops when the index reading covers a period of one or more revelations from the prover hand on the index. 6- The nozzles are indicated by a stamps which reads the amount of time required for the passage for one cubic foot of air at specific temperature and pressure. 7- Knowing the exact amount of air or gas passing through the nozzle at given temperature and pressure, the accuracy of the meter can be calculated. Results: After the experiment concluded, it has been found out that factor for air proving at room temperature was 101.7 using table 2. Also, Fr has been found using Table 3. Regardless the nozzle in-place, the air factor did not change much due to very close values. The accuracy of the diphgram was near 99% which is great. These calculations have many sources of error. One of which is timing accuracy. Also, calculations might been wrong due to change in the units, etc.… Finally, we assumed if we had our percent proof above 100% the nozzle was slow. Summer 2020 size code Vt cu.ft B 10 D 40 F 100 PNGE470 t ts sec sec 177.7 1.7 180.58 1.3 199.48 1.23 PN psig 29.7 28.2 26.7 PM psig 30.7 29.2 27.7 TN f 537 537 537 TM f 537 537 537 Fa Fr 101.7 1.002 101.7 1.001 101.7 1.001 %proof 99.7 99.87 99.54 Disscussion: In the beginning of the experiment, every nozzle was saturated and caliprated at a given pressure and temperature a cubic foot of air will pass through the orifice in a given amount of time. Since we have known amount of fluid passing through the nozzle, the accuracy can be calculated using the meter. For dry-air, the repeatability of the measurements is constant with that observed in previous measurements, with typical standard deviation at certain percentage. At higher humidity levels, the variation is slightly larger, which may be attributed to possible instability of the humidity level attained in the bell. It is really important to know that the effect due to the change in the density of the air due to the addition of water vapor are already included in the sonic nozzle prover equation and it is due to the change in the effective area of the nozzle. A variety of errors might occur in this experiment such as the nozzle not being manufactured to a certain specifications or by the ambient temperature being higher or lower than the calibration temperature. Errors can be also interduced by a faulty stopwatch or to start and stop the timer at wrong timing. Errors can be limited by limiting the amount of involvement in the lab. Summer 2020 Appendix: Table1 PNGE470 Summer 2020 Table 2 PNGE470 Summer 2020 Table 3 References: - Notes on ecampus. PNGE470 Purchase answer to see full attachment

Don't use plagiarized sources. Get Your Custom Essay on
PNGE 470 Colorado State University Accuracy of Positive Displacement Meters Lab Report Hi there I attached the instructions for the lab and 2 samples work,
Just from $13/Page
Order Essay
Homework On Time
Calculate the Price of your PAPER Now
Pages (550 words)
Approximate price: -

Why Choose Us

Top quality papers

We always make sure that writers follow all your instructions precisely. You can choose your academic level: high school, college/university or professional, and we will assign a writer who has a respective degree.

Professional academic writers

We have hired a team of professional writers experienced in academic and business writing. Most of them are native speakers and PhD holders able to take care of any assignment you need help with.

Free revisions

If you feel that we missed something, send the order for a free revision. You will have 10 days to send the order for revision after you receive the final paper. You can either do it on your own after signing in to your personal account or by contacting our support.

On-time delivery

All papers are always delivered on time. In case we need more time to master your paper, we may contact you regarding the deadline extension. In case you cannot provide us with more time, a 100% refund is guaranteed.

Original & confidential

We use several checkers to make sure that all papers you receive are plagiarism-free. Our editors carefully go through all in-text citations. We also promise full confidentiality in all our services.

24/7 Customer Support

Our support agents are available 24 hours a day 7 days a week and committed to providing you with the best customer experience. Get in touch whenever you need any assistance.

Try it now!

Calculate the price of your order

Total price:
$0.00

How it works?

Follow these simple steps to get your paper done

Place your order

Fill in the order form and provide all details of your assignment.

Proceed with the payment

Choose the payment system that suits you most.

Receive the final file

Once your paper is ready, we will email it to you.

Our Services

No need to work on your paper at night. Sleep tight, we will cover your back. We offer all kinds of writing services.

Essays

Essay Writing Service

You are welcome to choose your academic level and the type of your paper. Our academic experts will gladly help you with essays, case studies, research papers and other assignments.

Admissions

Admission help & business writing

You can be positive that we will be here 24/7 to help you get accepted to the Master’s program at the TOP-universities or help you get a well-paid position.

Reviews

Editing your paper

Our academic writers and editors will help you submit a well-structured and organized paper just on time. We will ensure that your final paper is of the highest quality and absolutely free of mistakes.

Reviews

Revising your paper

Our academic writers and editors will help you with unlimited number of revisions in case you need any customization of your academic papers